User Manual Thermo Fisher Scientific

User Manual Thermo Fisher Scientific-PDF Download

  • Date:30 Jul 2020
  • Views:2
  • Downloads:0
  • Pages:110
  • Size:1.75 MB

Share Pdf : User Manual Thermo Fisher Scientific

Download and Preview : User Manual Thermo Fisher Scientific


Report CopyRight/DMCA Form For : User Manual Thermo Fisher Scientific


Transcription:

Important Note, Please read this user guide thoroughly before using your meter Any use. outside of these instructions may invalidate your warranty and cause. permanent damage to the meter,Contact Information, For assistance with Thermo Scientific Orion products contact Technical Support by email at. wai techservbev thermofisher com or by phone within the United States call 1 800 225 1480. and outside the United States call 1 978 232 6000 or fax 1 978 232 6031. For additional product information contact your local authorized dealer local Thermo Scientific. Orion technical sales representative or contact us using the Water and Laboratory Products. WLP information on the page back of this manual,Applications and Technical Resources. Visit www thermoscientific com water to view Thermo Scientific Orion products and download. product literature user guides and manuals software updates and the latest application and. technical resources,Table of Contents,Chapter 1 6,Meter Introduction 6. Meter Overview 6,Packing List 7,Intended Use 7,Chapter 2 8.
Meter Basics 8,Using the Universal Power Adapter 8. Installing Batteries Optional Power Source 9,Attaching the Electrode Stand and Holder 10. Using the Electrode Holder 11,Meter Connections 12. Meter Keypad 14,Function Keys 15,Meter Display 16,Measurement Display Icons 18. pH Electrode Condition Icon 18,Meter Models and Measurement Capabilities 19.
Meter Maintenance 19,Chapter 3 20,Meter Setup Menus 20. Main Setup Menu 20,General Setup Menu Navigation 20. Channel Specific Method Mode and Temperature Menus 24. Method Menu 25,Mode and Settings Menu 28,Temperature Menu 41. Instrument Settings Setup Menu 45,Log View Menu 46. Data Log 46,Calibration Log 47,Diagnostics Menu 48.
Meter Self Test Procedure 48,Electrode Stability Test Procedure 49. Chapter 4 50,Using the pH or pH ISE Channel 50,Meter and Electrode Preparation 50. pH Calibration Procedure 51,pH Calibration Editing 52. ORP Calibration Procedure Relative mV Mode 53,ORP Calibration Procedure EH Units 54. ISE Calibration Procedure 55,ISE Calibration Editing 56.
Measurement Procedure 57,Chapter 5 58,Using the Conductivity Channel 58. Meter and Sensor Preparation 58,Conductivity Calibration Procedure 59. Conductivity Calibration Editing 60, Alternative Certified Cell Constant Entry Calibration Procedure 60. Conductivity Standards vs Temperature Table 61,Conductivity Meter Verification Procedure 62. Measurement Procedure 63,Chapter 6 64,Using the DO RDO Channel 64.
Meter and Sensor Preparation 64,Dissolved Oxygen Calibration Procedure 65. Air Water saturated Air Calibration 66,Water Air saturated Water Calibration 66. Manual Winkler Calibration 67,Set Zero Calibration 68. Measurement Procedure 69,Chapter 7 70,Data Transfer and Software Updates 70. Data Storage and Transfer Settings 70,Measurement Read Type Setting 70.
Data Log Setting 71,Export Data Settings 72,Printer Compatibility and Requirements 74. Computer Compatibility and Requirements 75,Using the USB to Serial Computer Cable 76. Using the USB Computer Cable 78,Orion Star Com Communication Software 81. Meter Interfacing with Other Computer Programs 82,Star A200 A300 Meter Remote Control Protocols 83. Meter Software Upgrade Procedure 90,Chapter 8 94,Customer Services 94.
Troubleshooting Tips 95,Meter Factory Reset Procedure 96. Meter User Reset Procedure 96,Notice of Compliance 97. WEEE Compliance 97,Declaration of Conformity 98,Meter Specifications 99. Ordering Information 106,Meter Accessories Electrodes and Solutions 108. Meter Introduction,Meter Overview, Thermo Scientific Orion Star A210 series benchtop meters feature an informative easy to.
read graphic display with backlight and onscreen calibration and setup instructions for intuitive. user friendly operation with minimum training The comprehensive keypad with menu specific. function keys and useful shortcut keys allows fast and efficient meter navigation and control. The multilanguage user interface allows customization of the meter in a variety of local. languages and new languages can be added through software updates. Measurements can be taken quickly and reliably with the onscreen reading stability indicator. and selectable read modes Auto Read timed or continuous with hold function The data log. collects up to 2000 measurement sets with time and date stamp and the non volatile meter. memory preserves all data even with the loss of power Use Thermo Scientific Orion Star. Com software to facilitate data transfer from the meter to a computer via USB or RS232. export data to an Excel spreadsheet or comma separated value file csv and print data to a. network or local printer, Utilize the Orion Star stirrer probe for meter powered and meter controlled sample stirring with. five selectable stirring speeds Use the stirrer probe with the included electrode stand to simplify. placement into and out of solutions and eliminate the need for magnetic stir bars and plates. The IP54 rated dust and splash resistant meter housing offers the flexibility to place the meter. where it is needed on the bench or wall mounted to save space For added site adaptability. power the meter using the included universal power adapter or optional four AA batteries Six. Orion Star A210 series benchtop meters are available to meet your exact measurement needs. 6 Orion Star A210 Series Benchtop Meter User Manual Thermo Scientific. Chapter 1 Meter Introduction,Orion Star A211 pH Benchtop Meter. Measure pH mV relative mV or ORP with temperature,Orion Star A212 Conductivity Meter. Measure conductivity TDS salinity or resistivity with temperature. Orion Star A213 RDO DO Meter, Measure dissolved oxygen as saturation or concentration with temperature using either. RDO optical or polarographic DO sensors,Orion Star A214 pH ISE Meter.
Measure ion concentration using an ion selective electrode ISE pH mV relative mV or ORP. with temperature,Orion Star A215 pH Conductivity Meter. Measure pH mV relative mV or ORP with temperature on channel one and. measure conductivity TDS salinity or resistivity with temperature on channel two. Orion Star A216 pH RDO DO Meter, Measure pH mV relative mV or ORP with temperature on channel one and. measure dissolved oxygen as saturation or concentration with temperature on channel two. Packing List, Orion Star A210 series benchtop meters and meter kits include the following items. Meter attachable electrode stand and holder,Universal power adapter. Literature CD,Printed quick start guide,Computer interface cable.
Meter test certificate, Each Orion Star A210 series meter is fully tested and certified by Thermo Fisher Scientific and. a certificate of calibration is included with each meter For specific meter and kit contents refer. to the Ordering Information section Visit our website at www thermoscientific com OrionMeters. to download the complimentary Orion Star Com data transfer computer software and Star. A200 A300 series USB driver,Intended Use, Please read this reference guide thoroughly Any use outside of these instructions may. invalidate the meter warranty and cause permanent damage to the meter. Thermo Scientific Orion Star A210 Series Benchtop Meter User Manual 7. Meter Basics,Using the Universal Power Adapter, A universal power adapter Catalog Number 1010003 with US EU UK and China plug plates. is included with the Orion Star A210 series benchtop meters This universal power adapter is. specifically for use with Star A210 series benchtop meters Use of other power adapters can. damage the meter and void the warranty, 1 Select the appropriate plug plate for the power outlet that will be used. 2 Remove the clear plastic cover from the groove on the back of the power adapter. 3 Slide the appropriate plug plate into the groove on the back of the power adapter. 4 Connect the assembled power adapter to a power outlet and the meter input labeled Power. A surge protector or uninterrupted power supply UPS is also recommended. 8 Orion Star A210 Series Benchtop Meter User Manual Thermo Scientific. Chapter 2 Meter Basics,Installing Batteries Optional Power Source.
1 Select four new AA alkaline batteries,2 Confirm that the meter is powered off. 3 Turn the meter over with the meter display facing down on a clean dry surface. 4 Remove the battery compartment cover push down on the battery compartment tab and. lift the battery compartment cover up, 5 Install the new batteries in the battery compartment orientating the batteries as shown in. the battery compartment housing, 6 Replace the battery compartment cover and turn the meter over with the meter display. Thermo Scientific Orion Star A210 Series Benchtop Meter User Manual 9. Chapter 2 Meter Basics,Attaching the Electrode Stand and Holder. The electrode stand can be attached to either side of the meter and up to two stands can be. attached to each meter A weighted base Catalog Number STARA HB is also available to. support the stand without attachment to the meter, 1 Open the box containing the electrode stand The box will include a base plate electrode.
arm pivot pin electrode holder ATC spacer and RDO clip. 2 Turn the meter over with the meter display facing down on a clean dry surface. 3 Identify the side of the meter that the stand will be installed on and remove the screw. between the circles on that side of the meter, 4 Align the base plate of the stand with the circles on the meter. 5 Replace the screw from step 3 to attach the base plate to the meter. 6 Turn the meter over with the meter display facing up. 7 Insert the electrode arm into the metal post on the base plate. 8 Connect the electrode holder to the electrode arm using the pivot pin. 10 Orion Star A210 Series Benchtop Meter User Manual Thermo Scientific. Chapter 2 Meter Basics,Using the Electrode Holder, Place electrodes in the stand for easy movement in and out of containers during calibration. sample measurement and storage Recommended electrode positions are shown below. 15 mm holder for conductivity and,dissolved oxygen sensors. Holder with 20 angle for gas,sensing ion selective electrodes. 12 mm flexible holder for pH ORP,and ion selective electrodes.
Stirrer probe holder,12 mm holder with optional ATC. probe insert,Optional RDO optical dissolved,oxygen sensor holder. Thermo Scientific Orion Star A210 Series Benchtop Meter User Manual 11. Chapter 2 Meter Basics,Meter Connections, The following diagram depicts all the possible meter connections available on the Orion Star. A210 series benchtop meters Some Star A210 series benchtop meters will have fewer. connections depending on the meter s measurement capabilities. Connector Function Applicable Meter Models,Connect the universal power adapter included with. Power All Star A210 series meters,meter to power the meter.
Connect an Orion Star stirrer probe Catalog Number. Stirrer 096019 for meter controlled and meter powered sample All Star A210 series meters. stirring with five selectable speeds, Connect the USB cable for bi directional data transfer. USB All Star A210 series meters, and communication via USB to a printer or computer. Connect the RS232 cable for bi directional data transfer. RS232 All Star A210 series meters, and communication via RS232 to a printer or computer. Connect a pH electrode ORP redox electrode or ion Star A211 Star A214. selective electrode ISE with BNC connector Star A215 Star A216 meters. Connect a half cell reference electrode with standard 2 5 Star A211 Star A214. mm pin tip connector Star A215 Star A216 meters, Connect an ATC temperature probe with 8 pin MiniDIN Star A211 Star A214. connector Star A216 meters, Connect a conductivity sensor or ATC temperature probe.
Star A212 Star A215 meters,with 8 pin MiniDIN connector. Ground Use to reduce interference generated by other equipment All Star A210 series meters. Connect a dissolved oxygen sensor either RDO optical. DO RDO Star A213 Star A216 meters,or polarographic with 9 pin MiniDIN connector. Orion Star A210 series benchtop meters are compatible with the same electrodes and sensors. used with Thermo Scientific Orion Versa Star meters and legacy Thermo Scientific Orion. Star and Star Plus meters This includes the Thermo Scientific Orion ROSS Ultra Triode. pH ATC electrodes Thermo Scientific Orion ROSS pH electrodes and Thermo Scientific. Orion DuraProbe conductivity sensors, 12 Orion Star A210 Series Benchtop Meter User Manual Thermo Scientific. Chapter 2 Meter Basics,Orion Star A211 pH Meter Connections. Orion Star A212 Conductivity Meter Connections,Orion Star A213 RDO DO Meter Connections.
Orion Star A214 pH ISE Meter Connections,Orion Star A215 pH Conductivity Meter Connections. Orion Star A216 pH RDO DO Meter Connections, Thermo Scientific Orion Star A210 Series Benchtop Meter User Manual 13. Chapter 2 Meter Basics,Meter Keypad, The Orion Star A210 series benchtop meter keypad includes menu specific function keys that. update on the display for fast and efficient meter operation and shortcut keys for easy. navigation to the measurement setup data log and calibration log menus. Key Icon and Name Function, Press the f1 f2 and f3 function keys to perform the action shown above. Thermo Scientific Orion Star A2 10 Series Electrochemistry Benchtop Meters 68X000441 Revision A June 2015 User Manual