Thermo Scientific Orion VERSA STAR RDO DO Temperature

Thermo Scientific Orion Versa Star Rdo Do Temperature-PDF Download

  • Date:30 Jul 2020
  • Views:0
  • Downloads:0
  • Pages:34
  • Size:785.13 KB

Share Pdf : Thermo Scientific Orion Versa Star Rdo Do Temperature

Download and Preview : Thermo Scientific Orion Versa Star Rdo Do Temperature


Report CopyRight/DMCA Form For : Thermo Scientific Orion Versa Star Rdo Do Temperature


Transcription:

Ross and the COIL trade dress are trademarks of Thermo Fisher Scientific Inc and. its subsidiaries, AQUAfast AQUASensors BOD AutoEZ ionplus KNIpHE LogR No Cal ORION perpHect. PerpHecT pHISA pHuture Pure Water Sage ROSS ROSS Ultra Sure Flow Titrator PLUS and. TURBO2 are registered trademarks of Thermo Fisher Scientific Inc and its subsidiaries. A All in One Aplus AUTO BAR AUTO CAL Auto ID AUTO READ AUTO STIR. Auto Test AutoTration CISA digital LogR DuraProbe EZ Startup ISEasy. Low Maintenance Triode Minimum Stir Requirement MSR NISS Optimum Results Orion Dual. Star Orion Star SAOB SMART AVERAGING SMART STABILITY Star LogR Star Navigator 21. Stat Face Triode are trademarks of Thermo Fisher Scientific Inc and its subsidiaries. Guaranteed Success and The Technical Edge are service marks of Thermo Fisher Scientific Inc. and its subsidiaries, 2011 Thermo Fisher Scientific Inc All rights reserved All trademarks are the property of. Thermo Fisher Scientific Inc and its subsidiaries, The specifications descriptions drawings ordering information and part numbers within this. document are subject to change without notice, This publication supersedes all previous publications on this subject. Thermo Scientific Orion VERSA STAR RDO DO Temperature Module. Table of Contents,Chapter 1 Introduction 1,Chapter 2 Module Overview 3.
Module Connections and Inputs 3,Module Maintenance 4. Display Information From Module 4,Chapter 3 Setup Menus 5. Setup Menus for Primary Measurements 5,Read Type Information 6. RDO Probe Information 6,Setup Menu for Temperature 7. Chapter 4 RDO DO Calibration 9,Calibration Methods 9.
Calibration Procedure 9,Chapter 5 Temperature Calibration 11. Chapter 6 Measurement 13,Chapter 7 Methods 15,Chapter 8 Data Storage and Review 17. Data Storage Settings 17,Measurement Read Type 17,Datalog Computer and Printer Export Settings 17. Datalog Review 17,Calibration Review 18,Chapter 9 Customer Services 19. Troubleshooting Guide 19,Assistance 19,Warranty and Registration 19.
WEEE Compliance 19,Declaration of Conformity 20,RDO DO Temperature Module Specifications 21. Ordering Information 22,Appendix Advanced Features 27. Dissolved Oxygen Barometric Pressure Compensation 27. Dissolved Oxygen Salinity Correction 27,This page intentionally left blank. Chapter 1 Introduction, Thank you for your purchase of the Thermo Scientific Orion VERSA STAR RDO DO Temperature Module. This module is for use with the Thermo Scientific Orion VERSA STAR Advanced Electrochemistry Benchtop Meter. Some key features of this module include, Works with and automatically recognizes Orion polarographic DO and RDO optical.
luminescence base probes, Polarographic DO probe measurement range of 0 to 90 mg L and 0 to 600 saturation. RDO probe measurement range of 0 to 50 mg L and 0 to 500 saturation. Readings compensate for temperature salinity and barometric pressure. Automatic and manual datalogging options, Quick calibration using water saturated air air saturated water Winkler titration or zero point. Please read this reference guide thoroughly Any use outside of these instructions may invalidate your warranty and. cause permanent damage to the meter, Thermo Scientific Orion VERSA STAR RDO DO Temperature Module 1. This page intentionally left blank,Chapter 2 Module Overview. To connect the modules to the meter and for specific information regarding the meter outputs please see the. Orion VERSA STAR Reference Guide The Reference Guide is on the CD included with the meter and also available. at www thermoscientific com water,Module Connections and Inputs.
1 Select either the RDO or Polarographic DO probe based on sample range and testing requirements. 2 Place selected probe in the meter s electrode arm If using the RDO probe attach the included adapter. 3 Attach the 9 pin mini DIN connection into the RDO DO module. 4 Attach the stirrer to the side of the meter and into the meter s electrode arm if needed for your. test method, 5 Set up area with deionized rinse water sample and other supplies. 6 Connect the meter for data collection Hyperterminal LIMS Star Printer etc if desired. 7 Power the Orion VERSA STAR meter stirrer and data collection device s. RDO DO Module,9 pin mini DIN connector for,RDO or polarographic DO probe. Thermo Scientific Orion VERSA STAR RDO DO Temperature Module 3. Module Maintenance, For routine maintenance dust and wipe the module with a damp cloth If necessary warm water or mild water based. detergent can be used Maintenance can be performed on a daily weekly or monthly basis as required by operating. environment Immediately remove any spilled substance using the proper cleaning procedure for that spill type. Display Information from Module,Example RDO Measurement Screen Single Channel. Date and time,Primary reading,Temperature,RDO cap life Barometric.
information pressure,changes to red,color at 30 days. Calibration,information,The information on the display can be customized. Enter the setup menu to make changes,1 Press Setup. 2 Use the arrow keys to highlight Inst Settings,press f3 select key. 3 Use the arrow key to highlight Display View,Press f3 select key.
4 Follow the onscreen prompts to choose what you,would like to see on the measurement screen. 4 Thermo Scientific Orion VERSA STAR RDO DO Temperature Module. Chapter 3 Setup Menus,Setup Menus for Primary Measurements. 1 Select the RDO DO channel by pressing the channel key then press f3 Setup. The meter detects the RDO or DO probe type automatically and selects mode menu. 2 Select the options for your testing,RDO DO Setup Menu. Default values are in bold in the table below,Probe Serial Number Input. Sample ID Off Manual Auto Increment,Dependent on measurement unit.
Resolution 0 1 default 1 for sat,0 01 default 0 1 for mg L. Stability Smart Fast Medium Slow,Averaging Off Automatic Smart. Read Type AutoRead Timed Continuous,Measurement Unit mg L. Pressure Compensation Auto Manual,Salinity Correction Auto Manual. Alarm Limit CalDue Set Point, For automatic salinity correction of RDO DO reading.
ensure oxygen probe and conductivity cell are in the same sample. Limit and set point alarm default set to off,Cal due alarm set to 12 hours. 3 Press the f1 Back button to step back to the main screen. Thermo Scientific Orion VERSA STAR RDO DO Temperature Module 5. Read Type Information, There are three main read types available as shown in the setup menus. Auto Read The meter will display the measurement as it stabilizes and lock and hold the. measurement when it is stable The AR icon will lock on the screen Once the measurement is locked. the meter will automatically export the measurement to the data log if the data log is enabled in. the setup menu and to a printer or computer if a printer or computer is connected to the meter and. enabled in the setup menu To take a new measurement press measure esc 1. Timed The meter will read and display a measurement at the set time interval that is programmed. by the operator There are two options, 1 By checking selecting the Interval box at the set time interval the meter will automatically. export the measurement to the data log if the data log is enabled in the setup menu and to a. printer or computer if a printer or computer is connected to the meter and enabled in the setup. menu If selected enter the time interval value in a hours minutes seconds format using the. numeric keypad and press f3 select key, 2 If the Interval box is unchecked not selected a single timed measurement will be taken. The meter will apply the same saving and exporting data conditions as with the set time interval. to record and export data but lock after one timed interval has expired Press measure esc 1. to reset the countdown clock and take a new reading. Continuous The meter will continuously measure and update the display This read type is useful. when performing an experiment that requires continuous measurements to be taken Press the. log print 0 key to export the measurement to the data log if the data log is enabled in the setup. menu and to a printer or computer if a printer or computer is connected to the meter and enabled in. RDO Probe Information, The RDO probe works with a unique optical cap Information exclusive to this cap such as.
Serial number,Remaining cap life,Date of manufacture. can be found in the setup menu when the RDO probe is connected and recognized as the measurement mode. 1 Verify RDO probe is connected,2 Press setup 3, 3 Use the arrows to highlight the channel with RDO if needed Press f3 Select key. 4 Use the right arrow to highlight RDO Electrode Info to review information. 6 Thermo Scientific Orion VERSA STAR RDO DO Temperature Module. Setup Menu for Temperature, All of the temperature measurement settings are found in Temperature Setup menu for each measurement. Temperature Unit C or F selection, Temperature Input and Type Ch 1 ATC Ch 2 ATC Ch 3 ATC Ch 4 ATC MTC. Ch 1 ATC only or All ATC allows all ATCs to be,Temperature Cal.
calibrated at once, This number will change to match the setup menu channel. Temperature probes can be set up in the measurement mode of the sample you are measuring Ch 1 ATC allows the. temperature probe to be used in more than one channel for temperature compensated measurements such as when. a pH electrode and a 2 cell conductivity cell are in the same measurement sample Temperature is a critical pH and. conductivity measurement as the measurement values require a temperature correction to determine the reference. value often 25 C in the America and 20 C in Europe. Thermo Scientific Orion VERSA STAR RDO DO Temperature Module 7. This page intentionally left blank, 8 Thermo Scientific Orion VERSA STAR RDO DO Temperature Module. Chapter 4 RDO DO Calibration, Polarographic DO probes only These probes need to be polarized prior to use It is continuously polarized when. connected to the meter If the probe is new or has not been connected to the meter connect the probe to the module. connected to the meter connect the meter to a power source and wait 30 minutes for polarization. Calibration Methods, Water Saturated Air Air The simplest and most accurate method this method uses the. calibration sleeve included with the DO probe Due to the inherent differences between water. saturated air and air saturated water 102 3 saturation will be displayed when the calibration. reading is stable using a polarographic DO probe For best accuracy the calibration temperature. should match the measuring temperature Moisten the sponge or absorbent cloth in the calibration. sleeve with distilled water and insert the probe into the sleeve without touching the. water saturated material, Air Saturated Water Water Use water that is 100 saturated with air Bubble air into a water.
sample for an extended period of time preferably overnight. Winkler Titration Manual This method uses a water sample with a known concentration. of dissolved oxygen and can be used to calibrate the DO probe to the value achieved by a Winkler. titration A manual calibration involves performing a Winkler titration and using that sample as a. calibration standard The oxygen level resulting from the titration is entered This correlates the. meter input to the Winkler titration Due to the possibility of titration errors this method is inherently. less accurate, Set Zero A zero point calibration requires oxygen free solution This method needs to be repeated. each time the probe is replaced or modified cap change with RDO probes membrane change or. fresh electrolyte with polarographic DO probes This calibration is generally not required unless. measurements will be taken below 10 saturation or 1 mg L An air or water calibration needs to be. performed prior to a zero point calibration,Calibration Procedure. 1 Press cal to enter calibration mode,2 Follow the screen prompts for calibration. 3 The calibration can be saved or printed when complete. 4 Information for the last calibration can be displayed on the main measurement screen Refer to this. reference guide Chapter 2 Module Overview Display Information section for more information. Thermo Scientific Orion VERSA STAR RDO DO Temperature Module 9. This page intentionally left blank, 10 Thermo Scientific Orion VERSA STAR RDO DO Temperature Module. Chapter 5 Temperature Calibration, The ATC temperature display has a relative accuracy of 0 1 C Temperature sensors built into RDO polarographic.
DO probes have varying temperature accuracies usually 0 5 C to 2 C Use this function only if necessary Since the. temperature offset calculated during the calibration is applied to all future temperature measurements recalibrate if a. different probe is used For temperature calibration the probe needs to be connected to the meter and the calibration. solution should have a known stable temperature It is recommended that two NIST traceable thermometers be used. to measure and verify solution temperature,1 Press f3 setup key. 2 Use the arrow keys to highlight the channel that the temperature probe is connected to. Press f3 select, 3 Use the arrow keys to highlight Temperature and press f3 select. 4 Use the arrow keys to highlight the appropriate Temperature Cal option and press f3 select. Automatic and manual datalogging options Quick calibration using water saturated air air saturated water Winkler titration or zero point Please read this reference guide thoroughly Any use outside of these instructions may invalidate your warranty and cause permanent damage to the meter This page intentionally left blank Thermo Scientific Orion VERSA STAR RDO DO Temperature Module