Thermo Scientific Orion VERSA STAR pH Temperature Module

Thermo Scientific Orion Versa Star Ph Temperature Module-PDF Download

  • Date:30 Jul 2020
  • Views:2
  • Downloads:0
  • Pages:36
  • Size:743.55 KB

Share Pdf : Thermo Scientific Orion Versa Star Ph Temperature Module

Download and Preview : Thermo Scientific Orion Versa Star Ph Temperature Module


Report CopyRight/DMCA Form For : Thermo Scientific Orion Versa Star Ph Temperature Module


Transcription:

Ross and the COIL trade dress are trademarks of Thermo Fisher Scientific Inc and. its subsidiaries, AQUAfast AQUASensors BOD AutoEZ ionplus KNIpHE LogR No Cal ORION perpHect. PerpHecT pHISA pHuture Pure Water Sage ROSS ROSS Ultra Sure Flow Titrator PLUS and. TURBO2 are registered trademarks of Thermo Fisher Scientific Inc and its subsidiaries. A All in One Aplus AUTO BAR AUTO CAL Auto ID AUTO READ AUTO STIR. Auto Test AutoTration CISA digital LogR DuraProbe EZ Startup ISEasy. Low Maintenance Triode Minimum Stir Requirement MSR NISS Optimum Results Orion Dual. Star Orion Star SAOB SMART AVERAGING SMART STABILITY Star LogR Star Navigator 21. Stat Face Triode are trademarks of Thermo Fisher Scientific Inc and its subsidiaries. Guaranteed Success and The Technical Edge are service marks of Thermo Fisher Scientific Inc. and its subsidiaries, 2011 Thermo Fisher Scientific Inc All rights reserved All trademarks are the property of. Thermo Fisher Scientific Inc and its subsidiaries, The specifications descriptions drawings ordering information and part numbers within this. document are subject to change without notice, This publication supersedes all previous publications on this subject. Thermo Scientific Orion VERSA STAR pH Temperature Module. Table of Contents,Chapter 1 Introduction 1,Chapter 2 Module Overview 3.
Module Connections and Inputs 3,Module Maintenance 3. Display Information From Module 4,Chapter 3 Setup Menus 5. Setup Menus for Primary Measurements 5,Read Type Information 6. Isopotential Value 6,Linear Regression Information 6. Setup Menu for Temperature 7,Chapter 4 pH RmV and ORP Calibration 9.
pH Calibration 9,RmV Calibration 9,ORP Calibration 10. Chapter 5 Temperature Calibration 11,Chapter 6 Measurement 13. Chapter 7 Methods 15,Chapter 8 Data Storage and Review 17. Data Storage Settings 17,Measurement Read Type 17,Datalog Computer and Printer Export Settings 17. Datalog Review 17,Calibration Review 18,Chapter 9 Customer Service 19.
Troubleshooting Guide 19,Assistance 19,Warranty and Registration 19. WEEE Compliance 19,Declaration of Conformity 20,pH Temperature Module Specifications 21. Ordering Information 22,Appendix Advanced Features 27. Automatic pH Buffer Recognition Feature 27, Testing a pH Electrode for Automatic Buffer Recognition 27. Electrode Condition Icon 28,Multipoint Calibration Curve Feature 29.
This page intentionally left blank,Chapter 1 Introduction. Thank you for your purchase of the Thermo Scientific Orion VERSA STAR pH Temperature Module This module is for. use with the Thermo Scientific Orion VERSA STAR Advanced Electrochemistry Benchtop Meter Some key features of. this module include, pH range from 2 to 20 with resolution options of 0 001 0 01 and 0 1. mV and RmV ranges from 2000 mV,Automatic and manual datalogging options. Up to 6 point pH calibration with automatic recognition for USA NIST and DIN buffers. The ability to fix calibration errors without a complete recalibration thanks to calibration editing. Please read this reference guide thoroughly Any use outside of these instructions may invalidate your warranty and. cause permanent damage to the meter, Thermo Scientific Orion VERSA STAR pH Temperature Module 1. This page intentionally left blank,Chapter 2 Module Overview.
To connect the modules to the meter and for specific information regarding the meter outputs please see the. Orion VERSA STAR Reference Guide The Reference Guide is on the CD included with the meter and also available. at www thermoscientific com water,Module Connections and Inputs. 1 Prepare the pH Triode or electrode ATC probe for use as instructed in user guide. 2 Place electrode and ATC and Orion Star stirrer probe cat no 096019 optional in meter s. electrode arm, 3 Attach the BNC and ATC connectors to the meter module If using the stirrer probe attach probe to. the meter and assign the probe to the channel in the setup menu for instrument settings. 4 Set up work area with calibration buffers rinse water sample other supplies. 5 Connect the meter for data collection Hyperterminal LIMS Star Printer etc if desired. 6 Power the Orion VERSA STAR meter Orion Star stirrer and data collection device s. 8 pin mini DIN for ATC,BNC connection with shorting cap. Reference pin socket,Module Maintenance, For routine maintenance dust and wipe the module with a damp cloth If necessary warm water or mild water based. detergent can be used Maintenance can be performed on a daily weekly or monthly basis as required by operating. environment Immediately remove any spilled substance using the proper cleaning procedure for that spill type. Thermo Scientific Orion VERSA STAR pH Temperature Module 3. Display Information from Module,Example pH Measurement Screen Single Channel.
Date and time,Primary reading Sample,temperature,Corresponding. Calibration information mV reading,with electrode slope Electrode condition. from last calibration and slope,and buffer points,The information on the display can be customized. Enter the setup menu to make changes,1 Press Setup. 2 Use the arrow keys to highlight Inst Settings,press f3 select key.
3 Use the arrow key to highlight Display View,Press f3 select key. 4 Follow the onscreen prompts to choose what you,would like to see on the measurement screen. Example Single Channel Display Screens for ORP RmV mV Modes. ORP screen RmV screen mV screen, 4 Thermo Scientific Orion VERSA STAR pH Temperature Module. Chapter 3 Setup Menus,Setup Menus for Primary Measurements. 1 Select the pH channel by pressing the channel key then press f3 Setup. 2 Scroll with the key highlight the appropriate pH module press f3 Select. 3 Arrow the cursor to Mode press f3 Select arrow to pH press f3 Select. 4 Select the options,pH Setup Menu,Default values are in bold in the table below.
Electrode ID Input,Sample ID Off Manual Auto Increment. Resolution 0 1 0 01 0 001,Stability Smart Fast Medium Slow. Averaging Off Automatic Smart,Read Type AutoRead Timed Continuous. Buffer Group USA DIN Custom,Isopotential Value Off On 7 000. Linear Regression Off On,Alarm Limit CalDue Set Point.
Limit and set point alarm default set to off,Cal due alarm set to 12 hours. ORP RmV and mV Setup Menus,Default values are in bold in the table below. Electrode ID Input,Sample ID Off Manual Auto Increment. Resolution 1 0 0 1,Stability Smart Fast Medium Slow. Averaging Off Automatic Smart,Read Type AutoRead Timed Continuous.
Alarm Limit CalDue Set Point, Note Different modes mV RmV ORP in the pH module have setup. options relevant to that mode Only for ORP and RmV modes. Limit and set point alarm default set to off,Cal due alarm set to 12 hours. 5 Press the f1 Back button to step back to the main screen. Thermo Scientific Orion VERSA STAR pH Temperature Module 5. Read Type Information, There are three main read types available as shown in the setup menus. Auto Read The meter will display the measurement as it stabilizes and lock and hold the. measurement when it is stable The AR icon will lock on the screen Once the measurement is locked. the meter will automatically export the measurement to the data log if the data log is enabled in. the setup menu and to a printer or computer if a printer or computer is connected to the meter and. enabled in the setup menu To take a new measurement press measure esc 1. Timed The meter will read and display a measurement at the set time interval that is programmed. by the operator There are two options, 1 By checking selecting the Interval box at the set time interval the meter will automatically. export the measurement to the data log if the data log is enabled in the setup menu and to. a printer or computer if a printer or computer is connected to the meter and enabled in the. setup menu If selected enter the time interval value in a hours minutes seconds format using. the numeric keypad and press f3 select key, 2 If the Interval box is unchecked not selected a single timed measurement will be taken.
The meter will apply the same saving and exporting data conditions as with the set time. interval to record and export data but lock after one timed reading has expired Press. measure esc 1 to reset the countdown clock and take a new reading. Continuous The meter will continuously measure and update the display This read type is useful. when performing an experiment that requires continuous measurements to be taken Press the. log print 0 key to export the measurement to the data log if the data log is enabled in the setup. menu and to a printer or computer if a printer or computer is connected to the meter and enabled in. Isopotential Value, The slope of all electrodes changes with temperature based on the Nerst equation The slope curves at the different. temperatures intersect at the isopotential point which is the concentration at which the potential of the electrode does. not vary with temperature For glass pH electrodes this value is 7 00. Linear Regression Information, The default is off which uses the traditional segmented approach The segmented approach is the most accurate and. precise More information can be found in the appendix section of this reference guide under the Multipoint Calibration. Curve Feature, Selecting on for this option results in the use of a single best fit line for the calibration curve and can be beneficial if. pH measurements vary greatly, 6 Thermo Scientific Orion VERSA STAR pH Temperature Module. Setup Menu for Temperature, All of the temperature measurement settings are found in Temperature Setup menu for each measurement.
Temperature Unit C or F selection, Temperature Input and Type Ch 1 ATC Ch 2 ATC Ch 3 ATC Ch 4 ATC MTC. Ch 1 ATC only or All ATC allows all ATCs to be,Temperature Cal. calibrated at once, This number will change to match the setup menu channel. Temperature probes can be set up in the measurement mode of the sample you are measuring Ch 1 ATC allows the. temperature probe to be used in more than one channel for temperature compensated measurements such as when. a pH electrode and a 2 cell conductivity cell are in the same measurement sample Temperature is a critical pH and. conductivity measurement as the measurement values require a temperature correction to determine the reference. value often 25 C in the America and 20 C in Europe. Thermo Scientific Orion VERSA STAR pH Temperature Module 7. This page intentionally left blank, 8 Thermo Scientific Orion VERSA STAR pH Temperature Module. Chapter 4 pH RmV and ORP Calibration,pH Calibration.
Up to six pH buffers can be used for calibration This can be a combination of automatically recognized USA NIST or. DIN buffer values and custom buffers Always use fresh pH buffers and select buffers that bracket the sample pH and. are at least one pH unit apart Prepare the pH electrode according to the instructions in the electrode user guide. 1 Press f1 Cal to perform a pH calibration, 2 Follow screen prompts for calibration Use fresh buffers rinse electrode with DI water before each. reading blot with a lint free tissue, 3 USA and DIN buffer sets are automatically recognized based off the selection in the setup menu. and the meter allows custom buffers to be used for up to 6 calibration points Custom buffer option. prompts for the nominal values and stores data, 4 During calibration points can be edited or retested prior to acceptance. 5 Information for the last calibration can be displayed on the main measurement screen Refer to this. reference guide Chapter 2 Module Overview Display Information section for more information. RmV Calibration, One standard can be used for relative mV calibration Always use fresh standard Prepare the electrode according to. the instructions in the electrode user guide,1 Press f1 Cal to enter calibration mode.
2 Follow the screen prompts for calibration, 3 The calibration can be saved or printed when complete. 4 Information for the last calibration can be displayed on the main measurement screen Refer to this. reference guide Chapter 2 Module Overview Display Information section for more information. Thermo Scientific Orion VERSA STAR pH Temperature Module 9. ORP Calibration, The mV reading of an ORP standard is dependent on the ORP electrode reference system reference material filling. solution and ORP standard temperature An ORP electrode can be automatically calibrated to read 420 mV relative. to the Standard Hydrogen Electrode E H at 25 C when the Orion ORP standard cat no 967901 or 96791. Orion ORP electrode cat no 9678BNWP 9778BNWP 9180BNMD or 9179BNMD and Orion 4M KCl filling solution. cat no 900011 or 4M KCl gel in electrode with cat no 9179BNMD are used The raw mV reading of an electrode. with the same conditions is 220 mV at 25 C, One standard can be used for ORP calibration Always use fresh standard Prepare the electrode according to the. instructions in the electrode user guide,1 Press f1 Cal to enter calibration mode. 2 Follow the screen prompts for calibration, 3 The calibration can be saved or printed when complete.
4 Information for the last calibration can be displayed on the main measurement screen Refer to this. reference guide Chapter 2 Module Overview Display Information section for more information. 10 Thermo Scientific Orion VERSA STAR pH Temperature Module. Power the Orion VERSA STAR meter Orion Star stirrer and data collection device s Module Maintenance For routine maintenance dust and wipe the module with a damp cloth If necessary warm water or mild water based detergent can be used Maintenance can be performed on a daily weekly or monthly basis as required by operating environment Immediately remove any spilled substance using the