Transcription:

Ross and the COIL trade dress are trademarks of Thermo Fisher Scientific Inc and. its subsidiaries, AQUAfast AQUASensors BOD AutoEZ ionplus KNIpHE LogR No Cal ORION perpHect. PerpHecT pHISA pHuture Pure Water Sage ROSS ROSS Ultra Sure Flow Titrator PLUS and. TURBO2 are registered trademarks of Thermo Fisher Scientific Inc and its subsidiaries. A All in One Aplus AUTO BAR AUTO CAL Auto ID AUTO READ AUTO STIR. Auto Test AutoTration CISA digital LogR DuraProbe EZ Startup ISEasy. Low Maintenance Triode Minimum Stir Requirement MSR NISS Optimum Results Orion Dual. Star Orion Star SAOB SMART AVERAGING SMART STABILITY Star LogR Star Navigator 21. Stat Face Triode are trademarks of Thermo Fisher Scientific Inc and its subsidiaries. Guaranteed Success and The Technical Edge are service marks of Thermo Fisher Scientific Inc. and its subsidiaries, 2011 Thermo Fisher Scientific Inc All rights reserved All trademarks are the property of. Thermo Fisher Scientific Inc and its subsidiaries, The specifications descriptions drawings ordering information and part numbers within this. document are subject to change without notice, This publication supersedes all previous publications on this subject. Thermo Scientific Orion VERSA STAR Advanced Electrochemistry Benchtop Meter. Table of Contents,Chapter 1 Introduction 1,Chapter 2 Meter Overview 3.
Meter Connections and Inputs 3,Connecting Electrode Arm 3. Meter Inputs 3,Connecting Power 4,Module Measurement Capabilities 4. Module to Meter Connections 4,Removing and Inserting Modules 5. Electrode Probe Attachments to the Module 6,Meter Display 7. Meter Keypad 8,Meter Maintenance 9,Chapter 3 Instruments Settings 11.
Chapter 4 Data Storage and Review 13,Data Storage Settings 13. Measurement Read Type 13,Computer and Printer Export Settings 13. Data Log Settings 13,Datalog Review 13,Calibration Review 14. Chapter 5 Customer Services 15,Troubleshooting Guide 15. Meter Diagnostics 15,Assistance 15,Warranty and Registration 15.
WEEE Compliance 15,Declaration of Conformity 16,Meter Specifications 17. Ordering Information 18,This page intentionally left blank. Chapter 1 Introduction, Thank you for your purchase of the Thermo Scientific Orion VERSA STAR Advanced Electrochemistry Benchtop Meter. This meter allows for up to four channels to be configured in any combination through its modular design Five modules. are available pH pH ISE conductivity dissolved oxygen or pH with LogR temperature technology All five modules. measure temperature and can be mixed and matched as needed and without the need to power off the meter. The large bright color display makes it easy to get the information and to use the advanced features Display. customization and information selection multilingual interface plain language prompts and easy to understand. graphics make this meter easy to use A RS232 communication port two USB communication ports and two. Orion Star stirrer ports are just some of the unique advantages of the Orion VERSA STAR. Please read this reference guide thoroughly Any use outside of these instructions may invalidate your warranty. and cause permanent damage to the meter, Thermo Scientific Orion VERSA STAR Advanced Electrochemistry Benchtop Meter 1. This page intentionally left blank,Chapter 2 Meter Overview.
Meter Connections and Inputs,Connecting Electrode Arm. The electrode arms can be attached to either or both sides of the meter A separate weighted base STARA HB. is sold separately to use the arm without attaching to the meter Most kits contain one arm and some kits provide. two arms A second arm may be attached on the other side of the meter through step 5 at the same time the first. arm is being attached,1 Open the box that contains the arm. 2 Turn the meter over on a clean dry surface, 3 Remove the screw between the two circles on the base plate. 4 Align the base plate with the raised circles on the meter. 5 Replace the screw removed in step 3 to attach the electrode. arm base to the bottom of the meter,6 Turn the meter over. 7 Insert the electrode arm over the metal post,Meter Inputs.
Orion VERSA STAR Meter Back Panel with No Modules Attached. Micro AB USB Mini B USB,Stirrer Stirrer,RS232 Power. Connector Description,Micro AB USB For data output connection. There are two stirrer jacks on the meter Each one accepts the Orion Star Stirrer Probe. cat no 096019 available in select meter kits or sold separately The probes can be used. Stirrer Jack, simultaneously and controlled separately by the meter through this connection For further. information on the stirrer probe assignments refer to chapter 3 instrument settings. RS232 Pin Use with the included RS232 cable for communications and data transfer via RS232 port. Use with the included mini USB to USB cable for communications and data transfer via USB. Mini B USB,Allows bidirectional communication, See the following section titled Connecting Power Removal of power cord from the connector. and reconnection reboots the meter, Ground Ground receptacle to reduce interference generated by other equipment.
Thermo Scientific Orion VERSA STAR Advanced Electrochemistry Benchtop Meter 3. Connecting Power, A universal power adapter with US EU UK and China plug plates is provided with the meter. 1 Select the appropriate plug plate for your power socket. 2 Slide the clear plastic cover off of the plug plate. 3 Slide the plug plate into the groove on the back of the power adapter. 4 Connect the power to the meter and to the power pin on the back of the meter. Note This adapter is specifically for use with the Orion VERSA STAR meter Use of other power adapters could cause. damage to the meter and will void any warranty, The module s connected to the meter define the meter parameters There are five available modules for use with the. four meter channels,Module Measurement Capabilities. The table below shows the module name label on the top tab of the module and measurement modes. All measurements include temperature,Module pH pH ISE RDO DO Conductivity pH LogR. Labeled As pH pH ISE RDO DO Cond pH LogR,Measurement RDO as.
pH ISE Conductivity pH,Modes Sat or mg L,DO as Temperature from pH. mV pH Resistivity,Sat or mg L bulb no ATC needed,RmV mV Temperature TDS mV. ORP Redox RmV Salinity RmV,Temperature ORP Redox Temperature ORP Redox. Temperature,Module to Meter Connections, Modules arrive installed in new Orion VERSA STAR meter and module packages. 1 Modules pH pH ISE Conductivity RDO DO each use one module slot. 2 pH LogR module requires two module slots,3 Open module slots are filled with blank modules.
Meter with blank pH LogR,and pH modules attached,Meter with blank Modules labeled for. RDO DO Conductivity easy identification,and pH ISE modules attached. 4 Thermo Scientific Orion VERSA STAR Advanced Electrochemistry Benchtop Meter. Removing and Inserting Modules, Easily change modules regardless of whether the meter is powered or unpowered. Module Removal, 1 With your thumb on the top labeled tab and your index finger on the bottom tab. squeeze the tabs towards each other, 2 Pull straight back to remove the module from the meter.
3 Powered meters will no longer display that channel s measurement. Module Insertion, 1 With your thumb on the top labeled tab and your index finger on the bottom tab. squeeze the tabs towards each other, 2 Push the module into an empty meter channel groove in the back of the meter. 3 Release the top and bottom tabs, 4 Powered meters will display channel information in top left of display. Moving the module in out of meter, Thermo Scientific Orion VERSA STAR Advanced Electrochemistry Benchtop Meter 5. Electrode Probe Attachments to the Module, The same electrodes and probes compatible with the Orion Star Star Plus and Star A series meters can be used with.
the Orion VERSA STAR meter and module system, 1 pH pH ISE electrodes connect to BNC ATC probe connects to top 8 pin mini DIN. 2 pH LogR electrode connects to BNC and temperature taken from the pH bulb. 3 Conductivity probes connect to the top 8 pin mini DIN. 4 RDO DO probes connect to the top 9 pin mini DIN and the probe type is automatically recognized. 1 pH and pH ISE Modules 2 pH LogR Module,3 Conductivity Module 4 RDO DO Module. For further information regarding each module please see the specific reference guide for that module. The Reference Guide is on the CD included with the meter and also available at www thermoscientific com water. 6 Thermo Scientific Orion VERSA STAR Advanced Electrochemistry Benchtop Meter. Meter Display, Pressing the channel key cycles through each individual channel When multiple modules are connected the channel. keys shows each possible combination of simultaneous channel displays To customize the channel key refer to the next. chapter chapter 3 Instrument Settings,Format View Contains Example Displays. Full screen Single Channel Ch Measurement Mode Method. 1 2 3 or 4 Main measurement display with units,Temperature units type.
Ready and Stability Indication,Sample ID Electrode ID. Last Calibration date time,cal points electrode slope. Half screen Channels 1 2 Ch Measurement Mode Method. 1 3 Main measurement display with units,1 4 Temperature units type. 2 3 Ready and Stability Indication,or 3 4 Sample ID Electrode ID. Last calibration data, Third of a screen Channels 1 2 3 Ch Measurement Mode Method.
1 2 4 Main measurement display with units,1 3 4 Temperature units type. or 2 3 4 Ready and Stability Indication,Last calibration data sample ID. Electrode ID, Fourth of a screen Channels 1 2 3 4 Ch Measurement Mode Method. Main measurement display,Temperature units type,Ready and Stability Indication. Last calibration data sample ID,Electrode ID, Secondary reading information shown per mode include mV for pH and ISE offset mV for ORP and RmV barometric.
pressure for RDO DO and cell constant for conductivity resistivity salinity and TDS For further information regarding. the exact information displayed for each module please see the specific reference guide for that module The. Reference Guide is on the CD included with the meter and also available at www thermoscientific com water. Thermo Scientific Orion VERSA STAR Advanced Electrochemistry Benchtop Meter 7. Meter Keypad, Press to perform the action shown above each key on the display. Press to switch the meter to standby mode with screensaver or to return to active. Press to view each individual channel and when multiple modules are connected each possible. combination of simultaneous channel displays To customize the channel key refer to the next chapter. chapter 3 Instrument Settings, In the measurement mode press begin a new measurement when in AutoRead mode AR icon. shown and locks reading on display when stable Press to use as an escape key in most meter menus. In data fields used to input the number 1, Press to scroll up through lists of items Also allows looping through the list goes to last item after. scrolling past the first item on the list In data fields used to input the number 2. From the measurement mode press to enter the setup menu In data fields used to input the number 3. Press to scroll to the left of displayed information In data fields used to input the number 4. Press to enter select options In data fields used to input the number 5. Press to scroll to the right of displayed information In data fields used to input the number 6. 8 Thermo Scientific Orion VERSA STAR Advanced Electrochemistry Benchtop Meter. Press to change the measurement mode of a channel In data fields used to input the number 7. Press to scroll down through lists of items Also allows looping through the list goes to first item after. scrolling past the last item on the list In data fields used to input the number 8. Press to view data and calibration logs In data fields used to input the number 9. Press to manually log print or log and print a measurement depending on your reading mode and. data output settings In data fields used to input the number 0. Press to enter calibration mode In data fields used to input a decimal point. Press to turn on and off stirrer probe in continuous or timed reading modes. Meter Maintenance, For routine meter maintenance dust and wipe the meter with a damp cloth If necessary warm water or mild water. based detergent can be used Meter maintenance can be performed on a daily weekly or monthly basis as required by. operating environment Immediately remove any spilled substance from the meter using the proper cleaning procedure. for that spill type, Thermo Scientific Orion VERSA STAR Advanced Electrochemistry Benchtop Meter 9.
This page intentionally left blank, 10 Thermo Scientific Orion VERSA STAR Advanced Electrochemistry Benchtop Meter. Chapter 3 Instrument Settings, From the main screen of the meter press f3 to enter the Instrument Setup menu The menus give the options. to personalize the meter for your lab The list of adjustable options are described in the table below. Instrument Settings Options,Level 1 Level 2 Level 3 Default settings in bold. Display Brightness 1 2 3 4 5,Display View Channel Temperature Sample ID. Measurement Mode Secondary Parameter Electrode Serial No. Method Calibration Details Default all,Stable User ID above set on.
Thermo Scientific Orion VERSA STAR Advanced Electrochemistry Benchtop Meter 1 Chapter 1 Introduction Thank you for your purchase of the Thermo Scientific Orion VERSA STAR Advanced Electrochemistry Benchtop Meter This meter allows for up to four channels to be configured in any combination through its modular design Five modules