Orion Versa Star Pro Electrochemistry Meters

Orion Versa Star Pro Electrochemistry Meters-PDF Download

  • Date:30 Jul 2020
  • Views:10
  • Downloads:0
  • Pages:146
  • Size:4.74 MB

Share Pdf : Orion Versa Star Pro Electrochemistry Meters

Download and Preview : Orion Versa Star Pro Electrochemistry Meters


Report CopyRight/DMCA Form For : Orion Versa Star Pro Electrochemistry Meters


Transcription:

Table of Contents,Chapter 1 6,Meter Introduction 6. Meter Overview 6,Packing List 7,Intended Use 7,Chapter 2 8. Meter Basics 8,Using the Universal Power Adapter 8. Attaching the Electrode Stand 9,Meter Connections 10. Meter Keypad 11,Function Keys 12,Numeric Keypad 12.
Measurement Modules 13,Module Measurement Capabilities 13. Preconfigured Meter and Module Options 13,Inserting and Removing Modules 14. Module Removal 14,Module Connections 15,Measurement Display Options 16. Single Dual or Multiple Channel Displays 16,Measurement Display Examples 18. Additional Measurement Display Icons 19,pH Electrode Condition Icon 20.
Customizing the Display 21,Meter and Module Maintenance 21. Chapter 3 22,Meter Setup Menus 22,Main Setup Menu 22. Channel 1 4 Setup Menus 24,Method Menu 26,Mode Menu 28. Temperature Menu 29,Instrument Settings Setup Menu 32. Log View Menu 34,Data Log 34,Calibration Log 35,Diagnostics Menu 36.
System Access Menu 37,Enabling the System Access Feature 38. Logging into the Administrator Account 42,Logging into Existing User Account 43. Logging Out of Existing User Account 43,Chapter 4 44. pH pH ISE and pH LogR Measurement Modules 44,Connecting Electrodes to the Module 44. Channel 1 4 Setup Menus 47,Mode Menus for pH mV Relative mV ORP and ISE 47.
pH Calibration 52,pH Calibration Editing 52,Relative mV Calibration 54. ORP Calibration 55,ISE Calibration pH ISE Module Only 56. ISE Calibration Editing 56,Temperature Calibration 57. ATC Probe Temperature Calibration 57,Manual Temperature Entry 58. LogR Temperature Calibration pH LogR Module Only 59. Incremental Techniques pH ISE Module Only 61,Single Known Addition 62.
Double Known Addition 64,Single Known Subtraction 66. Double Known Subtraction 68,Single Analate Addition 70. Single Analate Subtraction 72,Chapter 5 74,Conductivity Measurement Module 74. Connecting Electrodes to the Module 74,Channel 1 4 Setup Menus 75. Mode Menus for Conductivity TDS Salinity and Resistivity 75. Conductivity Calibration 80,Automatic or Direct Conductivity Calibration 80.
Cell Constant Entry Manual Conductivity Calibration 82. Conductivity Verification Procedure 83,Chapter 6 84. RDO DO Measurement Module 84,Connecting Electrodes to the Module 84. Channel 1 4 Setup Menus 85,Mode Menus for Dissolved Oxygen 85. RDO Electrode Information 86,Dissolved Oxygen Calibration 87. Water Saturated Air Calibration 87,Air Saturated Water Calibration 88.
Manual Winkler Calibration 88,Set Zero Calibration 89. Chapter 7 90,Meter Measurements 90,Measurement Overview 90. Measurement Read Types 92,Auto Read 92,Single Shot 93. Continuous 93,Measurement Procedure 94,Measurement Tip Using the Electrode Holder 95. Chapter 8 96,Data Transfer and Software Updates 96.
Data Transfer Settings 96,Installing the Versa Star USB Driver Software 98. Meter Software Upgrade Procedure 99,Computer and Printer Compatibility 101. Computer Requirements 101,Printer Requirements 101. Versa Star Pro Meter Remote Control Protocols 102,Remote Control Commands 102. Chapter 9 112,Customer Services 112,Troubleshooting Tips 113.
Meter Self Test 114,WEEE Compliance 115,Declaration of Conformity 115. Meter Specifications 116,Ordering Information 122,Versa Star Pro Meters and Versa Star Modules 122. Versa Star Pro Meter Accessories Electrodes and Solutions 126. Appendix A 128,Channel Specific Settings and Features 128. General Mode Settings 128,Sample ID 128,Stability 129. Averaging 129,Alarms 129,pH and ISE Mode Settings 130.
Buffer Group and Automatic Buffer Recognition 130,Isopotential Value 131. Linear Regression 132,Blank Correction 133,Low Level Stability 134. LogR Temperature Technology 135,LogR Temperature Accuracy 135. Advanced pH Electrode Diagnostics using LogR Technology 136. Conductivity TDS Salinity and Resistivity Mode Settings 140. Cell Constant K and Automatic Conductivity Calibration 140. Orion Conductivity Standard Values from 0 C to 50 C 141. Reference Temperature Temperature Compensation and Temperature Coefficient 143. Cell Type 144,TDS Factor 144,Salinity Type 144,Dissolved Oxygen Settings 145. Barometric Pressure Compensation 145,Salinity Correction 145.
Meter Introduction,Meter Overview, Thermo Scientific Orion Versa Star Pro meters offer four channels that can be configured. in any combination through their modular design Measurement modules are automatically. recognized by the meter and can be added or removed from channels at any time. Five Thermo Scientific Orion Versa Star measurement modules are available. VSTAR PH module measures pH mV relative mV or ORP with temperature. VSTAR ISE module measures concentration using an ion selective electrode ISE pH. mV relative mV or ORP with temperature, VSTAR LR module measures pH mV or ORP with temperature including LogR. technology that measures sample temperature using a glass bulb pH electrode no. separate ATC probe needed in samples, VSTAR CND module measures conductivity TDS salinity or resistivity with temperature. VSTAR RD module measures dissolved oxygen as percent saturation or concentration. with temperature using either RDO optical or polarographic probes. Orion Versa Star Pro meters offer a System Access feature for administrative and up to ten. user accounts with custom access to various meter functions including data time settings meter. setup measurement setup data log and calibration log Use password protection to secure. selected meter functions while leaving other functions accessible for all users the new System. Access feature allows you to customize meter access to meet your specific security and data. 6 Orion Versa Star Pro Meter User Manual Thermo Scientific. Chapter 1 Meter Introduction, reporting needs The system access feature can easily be turned on or off as the needs of your. lab change for the ultimate flexibility in operation and ease of use. Get the information you need quickly and easily from the large bright color display which can. be personalized to show only the measurements and information you need Display one to four. measurements simultaneously in any combination, When setting up the meter or performing a calibration step by step instruction prompts and a.
numeric keypad with menu specific function keys make meter operation fast and simple The. meter interface can be set to a variety of languages. Packing List, Versa Star Pro meters and meter kits include the following items. Meter attachable electrode stand,Universal power adapter 100 240 VAC 50 60 Hz. Literature CD,Printed quick start guide,USB computer cable. Meter test certificate, Each Versa Star Pro meter is tested and certified by Thermo Fisher Scientific and a certificate. of testing is included with each meter and measurement module system. For specific measurement module and kit contents refer to the Ordering Information section. Visit our website at www thermofisher com orionsoftware to download the latest Versa Star. USB driver,Intended Use, Please read this user manual thoroughly Any use outside of these instructions may invalidate.
the meter warranty and cause permanent damage to the meter. Thermo Scientific Orion Versa Star Pro Meter User Manual 7. Meter Basics,Using the Universal Power Adapter, A universal power adapter Catalog Number STARA PWR with US EU UK and China plug. plates is included with the Versa Star Pro meter, This universal power adapter is specifically for use with Versa Star Pro meters Use of other. power adapters can damage the meter and void the warranty The use of a surge protector or. uninterrupted power supply UPS is also recommended. 1 Select the appropriate plug plate for the power outlet. 2 Remove the clear plastic cover from the back of the power adapter. 3 Slide the plug plate into the groove on the back of the power adapter. 4 Connect the assembled power adapter to the outlet and meter input labeled POWER. 8 Orion Versa Star Pro Meter User Manual Thermo Scientific. Chapter 2 Meter Basics,Attaching the Electrode Stand. The electrode stand can be attached to either side of the meter and up to two stands can be. attached to each meter A weighted base Catalog Number STARA HB is also available to. support the stand without a meter, 1 Open the box containing the stand Included are base plate electrode arm pivot pin. electrode holder ATC spacer and RDO clip, 2 Turn the meter over with the meter display facing down on a clean dry surface.
3 Identify the side of the meter that the stand will be installed on and remove the screw. between the circles on that side of the meter, 4 Align the base plate of the stand with the circles on the meter. 5 Replace the screw from step 3 to attach the base plate to the meter. 6 Turn the meter over with the meter display facing up. 7 Insert the electrode arm into the metal post on the base plate. 8 Connect the electrode holder to the electrode arm using the pivot pin. 9 Once the stand has been assembled use the two wing nuts to adjust the tension at each. joint to restrict the movement forward and backward using the lower wing nut and upward. and downward using the upper wing nut,Cable Management Clips. ATC Spacer,Electrode Arm,Base Plate Lower,Electrode Holder Wing Nut. Thermo Scientific Orion Versa Star Pro Meter User Manual 9. Chapter 2 Meter Basics,Meter Connections, 1 Connect a Thermo Scientific Orion Star series stirrer probe Cat No 096019 to the. STIRRER 1 and or STIRRER 2 input on the meter The stirrer probes are controlled and. powered by the meter eliminating the need for magnetic stir bars and plates. The STIRRER 1 and STIRRER 2 inputs can be assigned to each channel and the. stirrer speed can be set using the Instrument Settings setup menu. 2 For data transfer to a computer connect a cable to the MINI B USB or RS232 input on the. meter and the corresponding input on the computer, 3 For data transfer to a printer connect a cable to the MICRO AB USB or RS232 input on the.
meter and the corresponding input on the printer, 4 Review the meter output settings Communication Printing Printing Format in the. Instrument Settings setup menu to ensure the meter output settings are correct to interface. the meter with a printer or computer, The following image shows the meter back panel without any modules connected. MICRO AB USB MINI B USB,STIRRER 1 STIRRER 2,RS232 POWER. Connector Function,MICRO AB USB Use for data output to a USB printer. STIRRER 1 Accepts an Orion Star stirrer probe which is controlled and powered by the meter. Use the RS232 cable Cat No 1010053 for bi directional communication and data. transfer via RS232, Use the USB cable included with meter for bi directional communication and data.
MINI B USB,transfer via USB, STIRRER 2 Accepts the Orion Star stirrer probe which is controlled and powered by the meter. Connect the universal power adapter included with meter to power the meter. Remove and reconnect the power adapter to reboot the meter. GND Ground receptacle used to reduce interference generated by other equipment. 10 Orion Versa Star Pro Meter User Manual Thermo Scientific. Chapter 2 Meter Basics,Meter Keypad, The Versa Star Pro meter keypad includes menu specific function keys that update on the. display for fast and efficient meter operation a numeric keypad to facilitate data entry and. shortcut keys for easy navigation to the measurement setup calibration data log and. calibration log menus,Key Icon and Name Function, Press the power key to enter the standby meter mode with screen saver. Press the power key a second time to enter the active meter mode. Press the f1 f2 and f3 function keys to perform the action shown above each. key on the display, Press the channel key when two or more modules are connected to the meter. to change the measurement display to show each channel individually or a. combination of two or more channels simultaneously. Press the measure esc key to start a new measurement when in the. measurement mode and using Auto Read or Single Shot read types. Press the measure esc key to escape from other menus and modes to the. measure esc measurement mode, Press to enter a value of one 1 when using the numeric keypad.
Press the up arrow key to scroll up through a list of items. Press to enter a value of two 2 when using the numeric keypad. Press the setup key to enter the setup menu from the measurement mode. Press to enter a value of three 3 when using the numeric keypad. Press the cal key to enter the calibration menu from the measurement mode. Press to enter a decimal point when using the numeric keypad. Thermo Scientific Orion Versa Star Pro Meter User Manual 11. Chapter 2 Meter Basics,Key Icon and Name Function, Press the left arrow key to scroll left through a list of items. Press to enter a value of four 4 when using the numeric keypad. left arrow,Press the enter key to select an option. Press to enter a value of five 5 when using the numeric keypad. Press the right arrow key to scroll right through a list of items. Press to enter a value of six 6 when using the numeric keypad. right arrow, Press the stirrer key to turn on or off the stirrer probe when in the. 8 Orion Versa Star Pro Meter User Manual Thermo Scientific 2 Meter Basics Using the Universal Power Adapter A universal power adapter Catalog Number STARA PWR with US EU UK and China plug plates is included with the Versa Star Pro meter This universal power adapter is specifically for use with Versa Star Pro meters Use of other